Waveform analysis
R&S
®
RTA4000
145
User Manual 1335.7898.02 ─ 08
"Enter"
The signal crosses the upper or lower level and thus enters the win-
dow made up of these two levels.
"Exit"
Searches for events when the signal leaves the window.
"Stay within"
The signal stays between the upper and lower level for a specified
time.
"Stay outside"
The signal stays above the upper level or below the lower level for a
specified time.
"Pass through" The signal crosses both levels in a specified time. The time measure-
ment starts when the signal crosses the first level.
"Not pass
through"
The signal crosses one level, but does not cross the second level in a
specified time.
Remote command:
Upper Level, Lower Level
Set the upper and lower voltage thresholds. When the signal crosses the first level, the
window measurement starts. It stops when the signal crosses the second level. To let
the instrument set the levels, tap "Find Threshold".
Remote command:
SEARch:TRIGger:LEVel:WINDow:LOWer
SEARch:TRIGger:LEVel:WINDow:UPPer
Polarity
Indicates the polarity of the pulse to be searched for.
Remote command:
SEARch:TRIGger:WINDow:POLarity
Comparison
Sets the condition how the measured pulse width is compared with the given limits.
The comparison works like the comparison of the width trigger, see
Remote command:
SEARch:TRIGger:WINDow:TIMerange
Search