![R&S RTA4000 Скачать руководство пользователя страница 69](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987069.webp)
Waveform setup
R&S
®
RTA4000
69
User Manual 1335.7898.02 ─ 08
"Auto"
The voltage range is set only at the oscilloscope by adjusting the ver-
tical scale.
"<High value>" Sets the higher voltage range of the connected probe. The selection
list shows the value that is specified on the probe.
"<Low value>"
Sets the lower voltage range of the connected probe. The selection
list shows the value that is specified on the probe.
Remote command:
Bandwidth Limit
Activates the lowpass filter in the probe control box. You can also set the filter directly
on the probe control box. The limit depends on the probe type and is shown on the
probe control box.
Remote command:
PROBe<m>:SETup:ADVanced:FILTer
Audible Overrange
Activates the acoustic overrange warning in the probe control box. You can also acti-
vate the sound directly on the probe control box.
Remote command:
PROBe<m>:SETup:ADVanced:AUDioverload
Info
Same as for active single-ended probes, see
5.4.3.5
Modular probes
The probes of the R&S
RT-ZM family are modular probes with multi-mode function.
They have a probe head and a probe amplifier connected by a cable, and various
probe tip modules and tip cables for different applications. R&S
RT-ZM probes are
equipped with Rohde
&
Schwarz probe interface, and provide special features: Probe-
Meter, micro button, and a wide offset compensation range.
Access: [Ch <n>] > "Probe" (scroll down).
Probes