![R&S RTA4000 Скачать руководство пользователя страница 330](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987330.webp)
Serial bus analysis
R&S
®
RTA4000
330
User Manual 1335.7898.02 ─ 08
Minimum/Maximum ← Data ← Cmd. and data
Opens the "Minimum"/"Maximum" submenu.
After setting the "Compare" condition, you can enter the value bit-by-bit by setting the
state high, low or do not care for each single bit. Alternatively, you can enter a hexa-
decimal value.
Remote command:
13.8.4
MIL-STD-1553 decode results
When the configuration of the serial bus is complete, the signal can be decoded:
1. In the "Bus" menu, enable "Decode".
2. In the "Display" menu, select the result display settings.
Chapter 13.1.2, "Displaying decode results"
3. In the "Bus Table" menu, enable the "Bus Table". Adjust the table settings.
Chapter 13.1.3, "Bus table: decode results"
The instrument captures and decodes the signal according to the protocol definition
and the configuration settings.
The color-coding of the various protocol sections and errors simplifies the interpretation
of the visual display. The decode information condenses or expands, depending on the
horizontal scale. Various data formats are available to show the result values.
Table 13-7: Content of the MIL-STD-1553 frame table
Column
Description
Start time
Time of word start in relation to the trigger point
Type
Word Type
RTA
RT address
Label
Symbolic label, available if a label list was loaded and applied
T/R
Data direction, transmit or receive
Sub
Subaddress
Length
Number of data bytes
Data
Hexadecimal values of the data bytes
RT/IMG
Response time/intermessage gap time
State
Overall state of the word
Remote commands are described in
Chapter 17.11.8.3, "MIL-1553 decode results"
MIL-STD-1553 (option R&S
RTA-K6)