![R&S RTA4000 Скачать руководство пользователя страница 558](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987558.webp)
Remote control commands
R&S
®
RTA4000
558
User Manual 1335.7898.02 ─ 08
YT
Default time diagram with a time axis in x-direction and the sig-
nal amplitudes displayed in y-direction.
XY
XY-diagram, combines the voltage levels of two waveforms in
one diagram.
*RST:
YT
DISPlay:XY:XSOurce
<Source>
Defines the source to be displayed in x-direction in an XY-diagram, replacing the usual
time base.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4
CH3 and CH4 are only available with 4-channel R&S
RTA4000
oscilloscopes.
*RST:
CH1
DISPlay:XY:Y1Source
<Source>
Defines the first source to be displayed in y-direction in an XY-diagram.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4
CH3 and CH4 are only available with 4-channel R&S
RTA4000
oscilloscopes.
*RST:
CH2
DISPlay:XY:Y2Source
<Source>
Defines an optional second source to be displayed in y-direction in an XY-diagram. The
command is only relevant for 4-channel R&S
RTA4000 instruments.
Parameters:
<Source>
NONE | CH1 | CH2 | CH3 | CH4
*RST:
NONE
17.8.6
Digital voltmeter
The DVM suffix <m> sets the number of the DVM measurement (measurement place).
............................................................................................ 559
DVM<m>:RESult[:ACTual]:STATus?
................................................................................ 560
Applications