![R&S RTA4000 Скачать руководство пользователя страница 657](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987657.webp)
Remote control commands
R&S
®
RTA4000
657
User Manual 1335.7898.02 ─ 08
TRIGger:A:LIN:IPERror
<IdParityError>
Triggers on a parity error. Parity bits are the bits 6 and 7 of the identifier. They verify
the correct transmission of the identifier.
is set to
ERRCondition
.
Parameters:
<IdParityError>
ON | OFF
*RST:
OFF
TRIGger:A:LIN:SYERror
<SyncError>
Triggers if synchronization caused an error.
is set to
ERRCondition
.
Parameters:
<SyncError>
ON | OFF
*RST:
OFF
TRIGger:A:LIN:ICONdition
<IdentifierCondition>
Sets the comparison condition for the identifier: If the pattern contains at least one X
(don't care), you can trigger on values equal or not equal to the specified value. If the
pattern contains only 0 and 1, you can also trigger on a range greater than or lower
than the specified value.
is set to
ID
or
IDDT
.
Parameters:
<IdentifierCondition> EQUal | NEQual | GTHan | LTHan
*RST:
EQ
TRIGger:A:LIN:IDENtifier
<Identifier>
Defines the identifier pattern.
is set to
ID
or
IDDT
.
Parameters:
<Identifier>
String containing binary pattern. Characters 0, 1, and X are
allowed. Enter the 6 bit identifier without parity bits, not the pro-
tected identifier.
TRIGger:A:LIN:DATA
<Data>
Defines the data pattern. The number of bytes in the data pattern is defined with
.
is set to
IDDT
.
Serial bus analysis