Remote control commands
R&S
®
RTA4000
596
User Manual 1335.7898.02 ─ 08
17.10.2
System settings
SYSTem:COMMunicate:INTerface[:SELect]
...................................................................... 597
....................................................................................597
...................................................................................... 598
........................................................................................ 598
......................................................................................... 598
................................................................................................598
.......................................................................................... 599
CALibration
Calibration
starts the self-alignment process. It can take several minutes. Consider
your timeout settings.
Calibration?
returns information on the state of the self-alignment. Return values ≠
0 indicate an error.
Same as
*CAL?
.
Return values:
<SelfAlignment>
Numeric status indicator
CALibration:STATe?
Returns the overall state of the self-alignment.
Return values:
<SelfAlignmentState> NOALignment | RUN | ERRor | OK | ABORt
NOALignment: no self-aligment was performed. Relevant for
service operations.
RUN: self-aligment is running
ERRor: an error occured.
OK: self-aligment has been performed successfully
ABORt: self-aligment has been cancelled
Usage:
Query only
SYSTem:NAME
Defines an instrument name.
General instrument setup