Remote control commands
R&S
®
RTA4000
621
User Manual 1335.7898.02 ─ 08
Parameters:
<Access>
READ | WRITe
*RST:
READ
TRIGger:A:I2C:AMODe
<AdrMode>
Sets the lenght of the slave address.
Parameters:
<AdrMode>
NORMal | EXTended
NORMal: 7 bit address
EXTended: 10 bit address
*RST:
NORMal
TRIGger:A:I2C:ADDRess
<AddressString>
Sets the address of the slave device. The address can have 7 bits or 10 bits.
Parameters:
<AddressString>
String with max. 7 or 10 characters, depending on the address
length. Characters 0, 1, and X are allowed, but X cannot be
assigned to a specified bit. If at least one X occurs in the
address, the complete address is set to X.
Example:
TRIG:A:I2C:AMOD NORM
TRIG:A:I2C:ADDR "1011"
TRIG:A:I2C:ADDR?
Return value (7bit address):
"0001011"
Example:
TRIG:A:I2C:AMOD EXT
TRIG:A:I2C:ADDR "10X1"
TRIG:A:I2C:ADDR?
Return value (10bit address):
"XXXXXXXXXX"
TRIGger:A:I2C:PATTern
<DataPattern>
Defines the bit pattern as trigger condition. Make sure that the correct pattern length
has been defined before with
Parameters:
<DataPattern>
String with max. 24 characters (3 byte * 8 bit). Characters 0, 1,
and X are allowed. X can be assigned to a specified bit. If you
define a pattern shorter than the pattern length, the missing LSB
are filled with X. If you define a pattern longer than the pattern
length, the pattern string is not valid
Example:
TRIG:A:I2C:PLEN 2
TRIG:A:I2C:PATT "10X10000XXXX1111"
TRIG:A:I2C:PATT?
Return value (2 bytes):
"10X10000XXXX1111"
Serial bus analysis