Power analysis (option R&S
RTA-K31)
R&S
®
RTA4000
382
User Manual 1335.7898.02 ─ 08
14.7.3.2
Configuring turn ON/OFF time measurements
For details of the configuration settings, see
Chapter 14.7.3.3, "Turn ON/OFF time set-
1. Press the Apps Selection key. Tap "Power Analysis".
2. In the "Power Path" tab, select "Turn ON/OFF time".
3. Connect the differential voltage probe and the current probe to the oscilloscope.
It is recommended that you use a high-voltage differential probe for measurements
at the DUT power input.
4. Demagnetize the current probe. For details, see the user manual of your current
probe.
5. Select the "Tools" tab.
6. Connect the probes to the R&S
RT-ZF20 power deskew fixture.
7. "Deskew" the probes and adjust the "Zero Offset".
8. Disconnect the probes from the R&S
RT-ZF20 power deskew fixture.
9. Connect the probes to the DUT:
● Connect the positive (+) signal socket of the differential probe to the line of the
AC input.
● Connect the negative (-) signal socket of the differential probe to the neutral of
the AC input.
● Connect the passive voltage probe to the DC output of the DUT.
● Connect the current probe to the input path of the load with the direction of the
arrow pointing towards the current flow.
10. Select the correct channels for the "Input Voltage" and the "Output Voltage" sour-
ces.
11. Select the "Type" for the measurement.
12. Press the Autoset button to adjust the display scales automatically or adjust them
manually.
On the screen, you can see the measurement waveform of the input and output
voltage. Also, the numeric measurement results are shown in the lower left corner.
For details, see
Chapter 14.7.3, "Turn ON/OFF time"
14.7.3.3
Turn ON/OFF time settings
Required probes:
●
Two voltage probes
●
Current probe
► To open the "Turn ON/OFF time" settings menu:
Power path power measurements