Remote control commands
R&S
®
RTA4000
716
User Manual 1335.7898.02 ─ 08
Suffix:
<n>
.
1..2
Only relevant if two current sources are used: 1 = Input, 2 = Out-
put
Parameters:
<CurrentSource>
CH1 | CH2 | CH3 | CH4 | RE1 | RE2 | RE3 | RE4
POWer:SOURce:VOLTage<n>
Sets the channel for the voltage source.
Suffix:
<n>
.
1..4
Only relevant if several voltage sources are used: 1 = In, 2 =
Out, 3 and 4 = Out for "Turn ON/OFF time" measurement
Parameters:
<VoltageSource>
CH1 | CH2 | CH3 | CH4 | RE1 | RE2 | RE3 | RE4
POWer:STATistics:RESet
Deletes the statistical results for the current measurement or all measurements,
respectivley, and starts a new statistical evaluation.
Usage:
Event
POWer:STATistics:VISible
Shows/ hides the statistical evaluation of the measurement.
Parameters:
<Visible>
ON | OFF
*RST:
OFF
17.12.2
Probe adjustment
POWer:DESKew[:EXECute]
Starts the automatic deskew procedure to align the waveforms of all visible channels. It
is necessary to deskew if a current and a voltage probe is used in the measurment.
Use the R&S
RT-ZF20 power deskew fixture to deskew the probes.
Usage:
Event
POWer:ZOFFset[:EXECute]
Executes a zero offset for all visible channels.
Usage:
Event
Power analysis (option R&S
RTA-K31)