Power analysis (option R&S
RTA-K31)
R&S
®
RTA4000
359
User Manual 1335.7898.02 ─ 08
Gate Count
Sets the number of inrush current gates. You can select up to three gates.
Remote command:
POWer:INRushcurrent:GATE<n>:STARt
POWer:INRushcurrent:GATE<n>:STOP
Set to Screen
Resets the cursors to their initial positions. This is helpful if the cursors have disap-
peared from the display or need to be moved for a larger distance.
14.5
Output power measurements
Output analysis is used to measure the characteristics of the output voltage.
............................................................................................... 365
14.5.1
Ripple
The " Ripple" analysis measures the ripple of the device output. In this measurement
the peak-to peak extremes of the output DC signal are of interest. The measurement
also includes the AC-RMS of the output DC signal, that is calculated as a standard der-
ivation.
14.5.1.1
Ripple results
The results of "Ripple" measurements are:
●
The voltage waveform
●
Numeric measurement results
Also, for each measurement result you can enable a statistic evaluation of the mea-
surement results. It returns the current, minimum and maximum measurement values,
the average and standard deviation, and the number of measured waveforms.
The numeric measurement results are defined as follows:
Table 14-3: Properties output ripple
Meas. type
Sym-
bol
Description/Result
Peak +
Vp+
Maximum value within the displayed section of the waveform.
Peak -
Vp-
Minimum value within the displayed section of the waveform.
Peak peak
Vpp
Peak-to-peak value of the waveform: the difference of maximum and mini-
mum values.
X
Ampl
= X
Max
- X
Min
Output power measurements