Serial bus analysis
R&S
®
RTA4000
301
User Manual 1335.7898.02 ─ 08
LIN Configuration Settings
Figure 13-31: LIN setup dialog
.......................................................................................... 301
Source
Sets the source of the data line. All channel waveforms can be used.
If the MSO option R&S
RTA-B1 is installed, you can use logic channels as source.
Remote command:
Idle
Defines the idle state of the bus. The idle state is the recessive state and corresponds
to a logic 1.
Remote command:
Threshold, Find Threshold
Set the signal threshold for the source channel. Enter a value, or use "Find Threshold"
to set the threshold to the middle reference level of the measured amplitude.
For analog channels, you can find the value also in the "Vertical" menu > "Chan-
nel
<n>" > "Threshold"
For logic channels, you can find the value also in the "Logic" menu > "Technology".
Remote command:
CHANnel<m>:THReshold:FINDlevel
Version
Selects the version of the LIN standard that is used in the DUT. The setting mainly
defines the checksum version used during decoding.
The most common version is LIN 2.x. For mixed networks, or if the standard is
unknown, set the LIN standard to "Any".
LIN (option R&S
RTA-K3)