![R&S RTA4000 Скачать руководство пользователя страница 498](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987498.webp)
Remote control commands
R&S
®
RTA4000
498
User Manual 1335.7898.02 ─ 08
NOR
"Not or" operator, none of the channels has the required state.
*RST:
AND
SEARch:TRIGger:PATTern:LEVel<n>
<ThresholdLevel>
Sets the threshold value for the specified source channel. You can set different levels
for the channels
Suffix:
<n>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ThresholdLevel>
Range:
Depends on vertical scale
SEARch:TRIGger:PATTern:LEVel<n>:DELTa
<LevelDelta>
Sets a hysteresis range to the threshold of the specified source channel to avoid
unwanted search results caused by noise oscillation around the level.
Suffix:
<n>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<LevelDelta>
Range:
Lower limit depends on vertical scale and other set-
tings, no upper limit
SEARch:TRIGger:PATTern:WIDTh:RANGe
<Range>
Sets the condition how the duration of a steady pattern is compared with the given ref-
erence time.
To set the reference value
width
, use
SEARch:TRIGger:PATTern:WIDTh[:WIDTh]
To set a range
Δt
SEARch:TRIGger:PATTern:WIDTh:DELTa
Parameters:
<Range>
WITHin | OUTSide | LONGer | SHORter
WITHin
Finds patterns steady for a time range
width ± Δt
.
OUTSide
Finds patterns outside a time range
width ± Δt
.
LONGer
Finds patterns steady for at least the given
width
.
SHORter
Finds patterns shorter than the given
width
.
Waveform analysis