![R&S RTA4000 Скачать руководство пользователя страница 488](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987488.webp)
Remote control commands
R&S
®
RTA4000
488
User Manual 1335.7898.02 ─ 08
17.6.4.1
General search settings
SEARch:STATe
<SearchState>
Enables and disables the search mode.
Parameters:
<SearchState>
ON | OFF
*RST:
OFF
Example:
Chapter 17.2.3.1, "Searching for a pulse of specified width"
SEARch:CONDition
<SearchCondition>
Parameters:
<SearchCondition>
EDGE | WIDTh | PEAK | RUNT | RTIMe | DATatoclock |
PATTern | PROTocol | WINDow
EDGE
An edge search result is found when the waveform passes the
given level in the specified direction.
WIDTH
A width search finds pulses with an exact pulse width, or pulses
shorter or longer than a given time, or pulses inside or outside
the allowable time range.
PEAK
The peak search finds pulses exceeding a given amplitude.
RUNT
The runt search finds pulses lower than normal in amplitude.
The amplitude crosses the first threshold twice without crossing
the second one. In addition to the threshold amplitudes, you can
define a time limit for the runt in the same way as for width
search: runts with exact width, shorter or longer than a given
time, or runts inside or outside the allowable time range.
RTIMe
The rise or fall time search finds slopes with an exact rise or fall
time, or rise/fall times shorter or longer than a given limit, or rise/
fall times inside or outside the allowable time range.
DATatoclock
The Data2Clock search - also known as setup/hold - finds viola-
tion of setup and hold times. It analyzes the relative timing
between two signals: a data signal and the synchronous clock
signal. Setup time is the time that the data signal is steady
before clock edge. Hold time is the time that the data signal is
steady after clock edge.
Waveform analysis