![R&S RTA4000 Скачать руководство пользователя страница 419](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987419.webp)
Signal generation (option R&S
RTA-B6)
R&S
®
RTA4000
419
User Manual 1335.7898.02 ─ 08
16.2.5
Settings for manual pattern
Access: "Menu" > "Pattern Gen." > "Pattern" = "Manual"
P0/P1/P2/P3
Sets the states to high or low for the respective pin of the manual pattern.
Remote command:
16.2.6
Settings for serial buses
You can use the pattern generator to generate signals according to serial protocols
(bus signals).
The genrated serial protocol signals are pseudo random pattern and are not adaptable.
Only the protocol type and the data rate can be selected.
The contact at the upper left is always ground and the signal levels are about 1 V. The
following table shows how the four outputs P0, P1, P2 and P3 are used, depending on
the signal.
Signal
P0
P1
P2
P3
UART
Tx
Rx
Unused
Unused
SPI
Clock
Mosi
Miso
Chip Select
I2C
Clock SCL
Data SDA
Unused
Unused
CAN
CAN H
CAN L
Unused
Unused
LIN
High
Low
Unused
Unused
Access: "Menu" > "Pattern Gen." > "Pattern" = "UART | SPI | I2C | CAN | LIN"
Pattern generator