TRIGger:A:HOLDoff:TIME
<HoldOffTime>
Defines the holdoff time. The next trigger occurs only after the holdoff time has passed.
Parameters:
<HoldOffTime>
Default unit: s
17.5.2
Edge trigger
............................................................................................... 462
..........................................................................................462
........................................................................................... 463
TRIGger:A:EDGE:FILTer:HFReject
.................................................................................. 463
................................................................................... 463
TRIGger:A:EDGE:SLOPe
<Slope>
Sets the slope for the edge trigger.
Parameters:
<Slope>
POSitive | NEGative | EITHer
POSitive
Rising edge, a positive voltage change
NEGative
Falling edge, a negative voltage change
EITHer
The rising as well as the falling edge
*RST:
POSitive
TRIGger:A:LEVel<n>[:VALue]
<Level>
Sets the trigger threshold voltage for edge, width, and timeout trigger.
Suffix:
<n>
.
1..5
Selects the trigger input. 1...4 select the corresponding analog
channel, 5 is the external trigger input. The number of channels
depends on the instrument.
Parameters:
<Level>
Range:
Depends on vertical scale.
Default unit: V
TRIGger:A:FINDlevel
Sets the trigger level to 50% of the signal amplitude.
Usage:
Event
Trigger