Remote control commands
R&S
®
RTA4000
463
User Manual 1335.7898.02 ─ 08
TRIGger:A:EDGE:COUPling
<Coupling>
Sets the coupling for the trigger source.
Parameters:
<Coupling>
DC | AC | LFReject
DC
Direct current coupling. The trigger signal remains unchanged.
AC
Alternating current coupling. A highpass filter removes the DC
offset voltage from the trigger signal.
LFReject
Sets the trigger coupling to high frequency. A 15
kHz highpass
filter removes lower frequencies from the trigger signal. Use this
mode only with very high frequency signals.
*RST:
DC
TRIGger:A:HYSTeresis
<Hysteresis>
Sets a hysteresis range around the trigger level. Hysteresis avoids unwanted trigger
events caused by noise oscillation around the trigger level. The automatic, small,
medium, large hysteresis values depend on the vertical scale.
Parameters:
<Hysteresis>
AUTO | SMALl | MEDium | LARGE
*RST:
AUTO
TRIGger:A:EDGE:FILTer:HFReject
<State>
Enables or disables an additional 5
kHz lowpass filter in the trigger path. This filter
removes higher frequencies and is available with AC and DC coupling.
To folter out higher frequencies, you can use either this command or
.
Parameters:
<State>
ON | OFF
*RST:
OFF
TRIGger:A:EDGE:FILTer:NREJect
<State>
Turns an additional 100 MHz lowpass filter in the trigger path on or off. This filter
removes higher frequencies and is available with AC and DC coupling.
To folter out higher frequencies, you can use either this command or
.
Trigger