![R&S RTA4000 Скачать руководство пользователя страница 775](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987775.webp)
Remote control commands
R&S
®
RTA4000
775
User Manual 1335.7898.02 ─ 08
17.13.2.1
Parallel bus - line configuration
BUS<b>:PARallel:WIDTh
<BusWidth>
Sets the number of lines to be analyzed for the parallel bus.
Suffix:
<b>
.
1..2
Selects the parallel bus.
Parameters:
<BusWidth>
Maximum number is the number of input channels.
Range:
1 to 4
Increment: 1
*RST:
4
Default unit: Bit
BUS<b>:CPARallel:WIDTh
<BusWidth>
Sets the number of lines to be analyzed for the parallel clocked bus.
Suffix:
<b>
.
1..2
Selects the parallel bus.
Parameters:
<BusWidth>
Range:
1 to 15 (clock only) or 14 (clock and CS)
Increment: 1
*RST:
4
Default unit: Bit
BUS<b>:PARallel:DATA<m>:SOURce
<DataSource>
BUS<b>:CPARallel:DATA<m>:SOURce
<DataSource>
Defines the logic channel that is assigned to the selected bit.
Use the command for each bit of the bus.
Suffix:
<b>
.
1..2
Selects the parallel bus.
<m>
Sets the bit number.
Parameters:
<DataSource>
D0..D15
Example:
BUS:PARallel:Width 4
BUS:PARallel:DATA0:SOURce D8
BUS:PARallel:DATA1:SOURce D9
BUS:PARallel:DATA2:SOURce D10
BUS:PARallel:DATA3:SOURce D11
Mixed signal option (option R&S