Serial bus analysis
R&S
®
RTA4000
335
User Manual 1335.7898.02 ─ 08
● "Transmission": on a transmission.
5. Set the other trigger parameters according to the signal characteristics. All settings
are described below.
ARINC 429 Trigger Settings
Figure 13-48: Trigger setup to trigger on label and data patterns
Word
Triggers on a word start or word stop.
Remote command:
Error
Triggers on any combination of protocol errors. Enable the error types that you want to
trigger on.
"Parity"
Checks the parity and triggers if the parity is even.
"Gap Error"
Triggers on a gap error. The gap is calculated automatically from the
set sample rate.
"Coding Error"
Triggers on a coding error.
Remote command:
Label
Set the label settings to be triggered on.
ARINC 429 (option R&S
RTA-K7)