Waveform setup
R&S
®
RTA4000
63
User Manual 1335.7898.02 ─ 08
5.4.3
Probe settings for probes with Rohde
&
Schwarz interface
Probes with Rohde
&
Schwarz probe interface have an integrated data memory that
contains identification data and individual probe correction parameters. The
R&S
RTA4000 can detect these probes and read out the data, for example, bandwidth,
termination and attenuation. These parameters do not need any adjustment.
Other parameters, like the offset value for the selected channel, the micro button action
and other probe-specific settings are defined in the probe menu. The probe menu
opens automatically when the probe is plugged in.
Access: [Ch <n>] > "Probe" (scroll down). Or: short channel menu > "Probe"
The menu provides the settings for the recognized probe. If no probe is recognized, the
default probe menu for BNC-connected probes is shown. The channel label at the bot-
tom shows the probe type, and coupling or termination.
5.4.3.1
Active broadband single-ended probes
Active single-ended probes like R&S
RT-ZS10 and R&S
RT-ZS20 provide special fea-
tures: the micro button and the ProbeMeter. The ProbeMeter measures DC voltages
between the probe tip and ground. Special features are available in the "Probe" menu
if the probe is connected and recognized by the instrument.
Access: [Ch <n>] > "Probe" (scroll down); or short menu > "Probe"
Figure 5-3: Probe settings of R&S
RT-ZS20 with ProbeMeter measurement
Offset
Same as "Offset" in the "Vertical" menu, see
Probes