![R&S RTA4000 Скачать руководство пользователя страница 341](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987341.webp)
Serial bus analysis
R&S
®
RTA4000
341
User Manual 1335.7898.02 ─ 08
SDI ← Label and Data
Sets the values for the source/destination identifier (SDI) bits.
Remote command:
Data ← Label and Data
Searches for a data, that is specified in the submenu.
Data Offset ← Data ← Label and Data
Sets a data offset.
Remote command:
SEARch:PROTocol:ARINc:DATA:OFFSet
Data Size ← Data ← Label and Data
Sets the data size.
Remote command:
SEARch:PROTocol:ARINc:DATA:SIZE
Compare ← Data ← Label and Data
Sets the condition how the decoded value is compared with the defined range.
Remote command:
SEARch:PROTocol:ARINc:DATA:CONDition
Data ← Data ← Label and Data
After setting the "Compare" condition, you can enter the value bit-by-bit by setting the
state high, low, or don't care for each single bit. Alternatively, you can enter a hexadeci-
mal value for each byte.
Remote command:
SEARch:PROTocol:ARINc:DATA:MAXimum
SEARch:PROTocol:ARINc:DATA:MINimum
13.9.6
ARINC 429 label list
Label lists are protocol-specific. An ARINC 429 file contains three values for each iden-
tifier:
●
"Arinc Label": the Arinc 429 label value, that identifies the data type and the param-
eters associated with it.
●
"Symbolic label": symbolic name of the label, specifying the device function.
●
"Word Format": an optional parameter to describe the composition of the ARINC
429 word. It can be provided as integer value or mnemonic.
Word format definition
0 or DATA = <P>< Data ><Label>
1 or DATA_SDI = <P>< Data ><SDI><Label>
2 or DATA_SSM = <P><SSM>< Data ><Label>
ARINC 429 (option R&S
RTA-K7)