5. Tap "Peak Setup".
6. Define the criteria for peak detection:
a) Set the "Min. Level". Magnitudes higher than this level are marked as peaks.
b) For a more detailed peak detection, enable the "Advanced Peak Setup".
Set the "Excursion", "Max. Width" and "Distance".
7. If necessary, set a reference marker:
a) Select the type of the "R-Marker".
b) If "R-Marker" = "Indicated", set the "Index".
c) If "R-Marker" ="Range", set the "Frequency" and "Span".
9.4.3.1
Marker menu
Access: "FFT" menu > enable "Peak List" > "Marker"
Source
Defines the waveform type that is searched for peaks.
Remote command:
Peak Setup
See
Min. Level
Defines the minimum absolute level for peak detection. See
R-Marker
See
Chapter 9.3.3.3, "R-Marker settings"
9.4.3.2
Peak setup
Access: "FFT" menu > enable "Peak List" > "Marker" > "Peak Setup"
The simple peak setup needs only the setting of the minimum level for peak detection.
More complex criteria can be defined in the advanced setup.
Spectrum analysis and spectrogram (option R&S