![R&S RTA4000 Скачать руководство пользователя страница 559](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987559.webp)
Remote control commands
R&S
®
RTA4000
559
User Manual 1335.7898.02 ─ 08
DVM<m>:ENABle
<VoltmeterEnable>
Enables and disables all configured meter measurements.
Suffix:
<m>
.
1..4
The suffix is irrelevant.
Parameters:
<VoltmeterEnable>
ON | OFF
*RST:
OFF
DVM<m>:SOURce
<Source>
Selects an analog channel as the source of the selected meter measurement.
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4
CH3 and CH4 are only available with 4-channel instruments.
DVM<m>:TYPE
<MeasurementType>
Sets the measurement type for the indicated DVM measurement.
Set
OFF
to disable the measurement.
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Parameters:
<MeasurementType> DC | ACDCrms | ACRMs | OFF
DC: mean value of the signal
ACDCrms: RMS value of the signal (AC+DC RMS)
ACRMs: RMS value of the signal's AC component (AC RMS)
OFF: disables the selected meter measurement.
*RST:
OFF
DVM<m>:RESult[:ACTual]?