Remote control commands
R&S
®
RTA4000
476
User Manual 1335.7898.02 ─ 08
TRIGger:EVENt:REFSave
<SaveReference>
If ON, the instrument saves reference waveforms of all active channels, and activates
the references. This action works only with single acquisition.
The channels are assigned to the references: C1 to R1, C2 to R2 and so on. If a chan-
nel is off, the assigned reference is also not active.
If "Nx
Single" > 1, the waveforms of the last acquisition are saved as reference wave-
forms.
Parameters:
<SaveReference>
ON | OFF
*RST:
OFF
TRIGger:EVENt:SCRSave
<SaveScreenshot>
If ON, the instrument saves a screenshot on each trigger event. The screenshot is
saved when the acquisition is complete. During saving, the acquisition stops, and
restarts when saving is finished. Thus, the waveform update rate decreases signifi-
cantly.
To set up the target directory, file name, color and file format, use the following com-
mands:
●
TRIGger:EVENt:SCRSave:DESTination
●
●
Parameters:
<SaveScreenshot>
ON | OFF
*RST:
OFF
TRIGger:EVENt:SCRSave:DESTination
<File>
Defines the target directory and file name for a screenshot that is saved on trigger
event when
is ON.
Parameters:
<File>
String parameter
String with path and file name
TRIGger:EVENt:TRIGgerout
<PulseOut>
If ON, the instrument generates a pulse on the Aux
Out connector on trigger event.
The acquisition is not delayed, the pulse generation runs asynchronously.
Enabling "Pulse" sets the "Setup" > "Aux Out" selection to "Trigger Out", and vice
versa. Selecting another item in the "Aux Out" menu disables the pulse action.
Trigger