Waveform setup
R&S
®
RTA4000
71
User Manual 1335.7898.02 ─ 08
"DM"
Differential mode input voltage (V
dm
), the voltage between the positive
and negative input terminal.
"CM"
Common mode input voltage (V
cm
), the mean voltage between the
positive and negative input terminal vs. ground.
"P"
Positive single-ended input voltage (V
p
). The voltage between the
positive input terminal and ground.
"N"
Negative single-ended input voltage (V
N
). The voltage between the
negative input terminal and ground.
Remote command:
Info
Same as for active single-ended probes, see
5.4.3.6
Power rail probes
R&S
RT-ZPR power rail probes have a Rohde
&
Schwarz probe interface with Probe-
Meter, but no micro button. R&S
RT-ZPR probes require 50
Ω
termination, which is set
automatically. The ProbeMeter measures DC voltages between the probe tip and
ground.
Access: [Ch <n>] > "Probe" (scroll down).
Figure 5-7: Probe settings of R&S
RT-ZPR20 with ProbeMeter measurement
Probes