![R&S RTA4000 Скачать руководство пользователя страница 76](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987076.webp)
Waveform setup
R&S
®
RTA4000
76
User Manual 1335.7898.02 ─ 08
"Sin(x)/x"
Two adjacent ADC sample points are connected by a sin(x)/x curve,
and also the adjoining sample points are considered by this curve.
The interpolated points are placed on the resulting curve. This inter-
polation method is the default method. It is precise and shows the
best signal curve.
"Linear"
Two adjacent ADC sample points are connected by a straight line, the
interpolated points are placed on the line. You see a polygonal wave-
form similar to the real signal, and also the ADC sample points as
vertexes.
"Sample-Hold"
The ADC sample points are displayed like a histogram. For each
sample interval, the voltage is taken from the sample point and con-
sidered as constant, and the intervals are connected with vertical
lines. Thus, you see the discrete values of the ADC.
Remote command:
Acquisition setup