![R&S RTA4000 Скачать руководство пользователя страница 279](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987279.webp)
Serial bus analysis
R&S
®
RTA4000
279
User Manual 1335.7898.02 ─ 08
UART configuration settings
Figure 13-18: UART setup dialog
.......................................................................................... 279
TX / RX / Source
Select the input channel of the UART lines. Input channels are the receive line (RX),
and the optional transmit line (TX).
If the MSO option R&S
RTA-B1 is installed, you can use logic channels as source.
Note:
TX is available only on bus 1 and bus 3. On bus 2 and bus 4, you can set only
one the RX line. The TX line occupies a second bus line. Thus, if TX is used on bus 1
or bus 3 , bus 2 or bus 4 is not available respectively.
Remote command:
Polarity
Selects if the transmitted data is high active (high = 1) or low active (low = 1). If RX and
TX are used, the setting affects both lines.
High active is used, for example, for control signals, while low active is defined for data
lines (RS-232).
Remote command:
Threshold, Find Threshold
Set the signal threshold for the source channel. Enter a value, or use "Find Threshold"
to set the threshold to the middle reference level of the measured amplitude.
UART / RS232 (option R&S
RTA-K2)