![R&S RTA4000 Скачать руководство пользователя страница 630](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987630.webp)
Remote control commands
R&S
®
RTA4000
630
User Manual 1335.7898.02 ─ 08
BUS<b>:UART:RX:SOURce
<Source>
BUS<b>:UART:DATA:SOURce
<Source>
Selects the input channel of the data line.
Suffix:
<b>
.
1..4
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4 | D0..D15
CH3 and CH4 are only available with 4-channel R&S
RTA4000
oscilloscopes.
*RST:
CH1
BUS<b>:UART:TX:SOURce
<Source>
Selects the input channel of the transmitter TX line.
Suffix:
<b>
.
1 | 3. Bus 2 or 4 is not available if the TX source is used on bus
1 or 3, respectively.
Parameters:
<TxSource>
CH1 | CH2 | CH3 | CH4 | NONE | D0..D15
NONE
Disables the optional TX line.
*RST:
NONE
BUS<b>:UART:POLarity
<IdleState>
Defines the logic levels of the bus. The idle state corresponds to a logic 1, and the start
bit to a logic 0. If both RX and TX lines are used, the setting affects both lines.
Alternative command for
Suffix:
<b>
.
1..4
Parameters:
<IdleState>
IDLLow | IDLHigh
IDLLow: idle low, low = 1
IDLHigh: idle high, high = 1
*RST:
IDLH
BUS<b>:UART:DATA:POLarity
<Polarity>
Defines if the transmitted data on the bus is high (high = 1) or low (low = 1) active. If
both RX and TX lines are used, the setting affects both lines.
Alternative command for
Serial bus analysis