Remote control commands
R&S
®
RTA4000
773
User Manual 1335.7898.02 ─ 08
LOGic<p>:DATA?
DIGital<m>:DATA?
Returns the data of the specified logic channel/pod for transmission from the instru-
ment to the controlling computer. The waveforms data can be used in MATLAB, for
example.
.
To set the range of samples to be returned, use
Suffix:
<m>
.
0..15
Selects the logic channel
<p>
1..2
Selects the logic pod
Parameters:
<WaveformData>
List of values according to the format settings.
Example:
FORM ASC,0
DIG1:DATA?
1,1,1,1,1,1,0,0,0,0,0,0,...
Usage:
Query only
LOGic<p>:DATA:HEADer?
DIGital<m>:DATA:HEADer?
Returns information on the waveform of the specified logic channel/pod.
Table 17-6: Header data
Position
Meaning
Example
1
XStart in s
-9.477E-008 = - 94,77 ns
2
XStop in s
9.477E-008 = 94,77 ns
3
Record length of the waveform in Samples
200000
4
Number of values per sample interval, usually 1.
1
Suffix:
<m>
.
0..15
Selects the logic channel
<p>
1..2
Selects the logic pod
Parameters:
<Header>
Comma-separated value list
Example:
-9.477E-008,9.477E-008,200000,1
Usage:
Query only
Mixed signal option (option R&S
RTA-B1)