![R&S RTA4000 Скачать руководство пользователя страница 455](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987455.webp)
Remote control commands
R&S
®
RTA4000
455
User Manual 1335.7898.02 ─ 08
If you set a specific value with
ment of a record length is turned off.
Parameters:
<AutoRecordLength> ON | OFF
Example:
ACQ:POIN:AUT ON
TIM:SCAL 1e-9
ACQ:POIN?;:SYST:ERR:ALL?
-> received 10000;0,"No error"
TIM:SCAL 5e-3
ACQ:POIN?;:SYST:ERR:ALL?
-> received 20000000;0,"No error"
ACQuire:POINts[:VALue]
Defines a record length value, the number of recorded waveform points in a segment.
The command turns
OFF.
If
ACQuire:POINts?
returns
the automatically set record length.
Each predefined record length corresponds to a maximum number of history seg-
ments, which are stored in the instrument's memory.Using the history, you can display
the history segments.
Available record length values are:
●
5 | 10 | 20 | 50 | 100 | 200 | 500 kSample
●
1 | 2 | 5 | 10 | 20 | 50 | 100 | 200 MSample
Parameters:
<RecordLength>
Record length in Samples.
If the entered value differs from the predefined values, the
instrument sets the closest value.
CHANnel<m>:TYPE
<DecimationMode>
Selects the method to reduce the data stream of the ADC to a stream of waveform
points with lower sample rate.
Suffix:
<m>
.
1..4
The command affects all channels regardless of the indicated
channel number. The suffix can be omitted.
Parameters:
<DecimationMode>
SAMPle | PDETect | HRESolution
SAMPle
Input data is acquired with a sample rate which is aligned to the
time base (horizontal scale) and the record length.
Waveform setup