User Manual 1335.7898.02 ─ 08
Parameters:
<AcquisitionTime>
Range and increment depend on time base (horizontal scale)
and other settings
Range:
250e-12 to 500
Increment: 1e-12
Default unit: s
TIMebase:DIVisions?
Queries the number of horizontal divisions on the screen.
Return values:
<HorizDivCount>
Returns 12 divisions.
Usage:
Query only
TIMebase:RATime?
Queries the real acquisition time used in the hardware. If FFT analysis is performed,
the value can differ from the adjusted acquisition time (
).
Return values:
<HWAcqTime>
Range:
Depends on various settings
Default unit: s
Usage:
Query only
17.4.8
Acquisition settings
............................................................................................454
................................................................................................ 455
.............................................................................................456
................................................................................................457
..............................................................................................457
...............................................................................................457
....................................................................................... 457
............................................................................................458
................................................................................................459
ACQuire:POINts:AUTomatic
<AutoRecordLength>
Enables or disables the automatic record length. The instrument sets a value that fits
to the selected timebase.
Waveform setup