![R&S RTA4000 Скачать руководство пользователя страница 457](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987457.webp)
Remote control commands
R&S
®
RTA4000
457
User Manual 1335.7898.02 ─ 08
ACQuire:PEAKdetect
<PeakDetect>
Enables or disables the peak detect acquisition mode.
You can use this command alternatively to
Parameters:
<PeakDetect>
AUTO | OFF
*RST:
OFF
ACQuire:HRESolution
<HighRes>
Enables or disables the high resolution acquisition mode.
You can use this command alternatively to
Parameters:
<HighRes>
AUTO | OFF
*RST:
OFF
ACQuire:NSINgle:COUNt
<NSingleCount>
Sets the number of waveforms acquired with
.
Parameters:
<NSingleCount>
Number of waveforms.
Range:
1 to maximum number that depends on the record
length.
*RST:
1
ACQuire:AVERage:COUNt
Defines the number of waveforms used to calculate the average waveform. The higher
the number, the better the noise is reduced.
Parameters:
<AverageCount>
Range:
2 to 100,000
*RST:
2
ACQuire:AVERage:RESet
Deletes the waveform and restarts the average calculation.
Usage:
Event
ACQuire:AVERage:COMPlete?
Returns the state of averaging.
Return values:
<AverageComplete> 0 | 1
Waveform setup