2-6 Construction of System
The following figure shows the construction sample of the system.
ABCD
ABCD
ABCD
BCD
a
u
w
y
q
i
o
t
r
e
1!
No.
Description
Installation and control sample
q
Laser marker head
-
w
Laser marker controller
-
e
PC for laser marker setting/monitoring
-
r
Protective enclosure
-
t
Laser protection shutter for work piece
gateway
Construct a control system which will separate (cut off) the laser beams
or shut off the laser power when it is opened.
y
Emergency stop button
Construct a control system for shutting off the laser power source when
it is opened.
u
Safety switch
i
Door for the maintenance
o
Safety relay unit or safety PLC, etc.
Connect both INTERLOCK 1 (X16, X17) and INTERLOCK 2 (X18,
X19) of the laser marker to the devices
t
to
i
using the relay output
terminal of the safety relay unit or the safety PLC.
Besides INTERLOCK terminals, it can also be connected to other I/O signals
such as LASER STOP terminals with the devices
t
to
i
depending on
the control specifications.
1)
Laser radiation warning light
To indicate the laser marker status, connect the output signals such as
ALARM OUT (Y15) or LASING OUT (Y20) to the light.
1!
Dust exhauster
Place the intake vent of the exhauster as close as possible to the
marking position of the work pieces.
ワㄐㄕㄊㄆ
• INTERLOCK terminals (X16, X17 and X18, X19) are connected to the operating coil of the internal relay in the controller.
Connect INTERLOCK (+) and INTERLOCK (-) in the I/O terminal with the non-voltage contact (dry contact) such as a
relay or a switch. Do not connect with the voltage contact such as a transistor.
ンㄆㄇㄆㄓㄆㄏㄆ
• When constructing a system compatible to ISO11553-1/ISO13849-1, construct the redundant interlock system by using
both INTERLOCK 1 (X16, X17) and INTERLOCK 2 (X18, X19) connected to safety PLC or safety relay unit to shut off the
laser power supply in accordance with the standards.
ME-LPGS-SM-8
Содержание LP-GS Series
Страница 17: ...1 Product Overview ME LPGS SM 8 ...
Страница 32: ...2 Laser Marker Installation ME LPGS SM 8 ...
Страница 48: ...3 Operation Method ME LPGS SM 8 ...
Страница 68: ...4 External Control Using I O ME LPGS SM 8 ...
Страница 107: ...5 External Control by Communication Commands ME LPGS SM 8 ...
Страница 116: ...6 Link Control with External Devices ME LPGS SM 8 ...
Страница 141: ...7 Maintenance ME LPGS SM 8 ...
Страница 161: ...161 MEMO ME LPGS SM 8 ...
Страница 162: ...Troubleshooting ME LPGS SM 8 ...
Страница 186: ...Index ME LPGS SM 8 ...
Страница 188: ...188 Unit power cable 26 USB 30 45 V Vibration resistance 33 W Warning 177 Z Z axis adjustment module 154 ME LPGS SM 8 ...
Страница 189: ......
Страница 190: ... Panasonic Industrial Devices SUNX Co Ltd 2014 2019 April 2019 ...