C h a p t e r 1 0 A L A R M S
7/26/00
INTEGRA Operations Revision A
10-69
20-0200-231 (CR), 20-0200-232 (Std), 20-0200-251 (CD)
Reason:
This alarm is only valid on systems with the Panel Door
Sensors and associated software installed. The alarm is posted if
the system detects that one of the system Panel Door has been
opened.
Disposition:
The system is put into STOP mode and an alarm is
posted.
GEM ALID: 146
GEM Alarm.
A
LARM
ID: 147
Description:
Input Station Exhaust Failure
Posting Program:
Monitor
Alarm Type:
Alarm & Sonalert
Reason:
This alarm is only valid on systems with the Input
Station Exhaust Sensor and associated software installed. The
Input Station Exhaust Sensor detects when no exhaust is present.
Disposition:
The system is put into CLEARING mode.
GEM ALID: 147
GEM Alarm.
A
LARM
ID: 148
Description:
Spin Station Exhaust Failure
Posting Program:
Monitor
Alarm Type:
Alarm & Sonalert
Reason:
This alarm is only valid on systems with the Spin Station
Exhaust Sensor and associated software installed. The Spin
Station Exhaust Sensor detects when no exhaust is present.
Disposition:
The system is put into CLEARING mode.
GEM ALID: 148
GEM Alarm.
A
LARM
ID: 149
Description:
Cabinet (System) Exhaust Failure
Posting Program:
Monitor
Alarm Type:
Alarm & Sonalert
ALCD
ALTX
ON CEID OFF CEID
6
146 Panel Door Sensors detects Panel Door open.
493
494
ALCD
ALTX
ON CEID OFF CEID
6
147 Input Station Exhaust failure.
495
496
ALCD
ALTX
ON CEID OFF CEID
6
148 Spin Station Exhaust failure
.
497
498
Содержание Synergy Integra
Страница 8: ......
Страница 24: ...1 4 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...
Страница 38: ...2 14 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...
Страница 222: ...9 62 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...
Страница 314: ...10 76 INTEGRA Operations Revision A 7 26 00 20 0200 231 CR 20 0200 232 Std 20 0200 251 CD...