2606B System SourceMeter® Instrument Reference Manual
Section 2: General operation
2606B-901-01 Rev. B / May 2018
2-69
Remote programming output-off states quick reference
The content of the following table is a quick reference of commands for programming output-off states
from a remote interface.
Output-off state programming quick reference
Command*
Description
smu
X
.source.offmode = smua.OUTPUT_NORMAL
Selects normal output-off mode.
smu
X
.source.offmode = smua.OUTPUT_HIGH_Z
Selects high-impedance output-off
mode.
smu
X
.source.offmode = smua.OUTPUT_ZERO
Selects zero output-off mode.
smu
X
.source.offfunc = smua.OUTPUT_DCVOLTS
Sets 0 V output with current limit
specified by the
smua.source.offlimiti
attribute.
smu
X
.source.offfunc = smua.OUTPUT_DCAMPS
Sets 0 A output with voltage limit
specified by the
smua.source.offlimitv
attribute.
smu
X
.source.offlimiti =
iValue
Sets current limit in normal
output-off mode.
smu
X
.source.offlimitv =
vValue
Sets voltage limit in normal
output-off mode.
* smu
X
can be
smua
for SMU Channel A or
smub
for SMU Channel B
USB storage overview
The 2606B includes a USB port on the front panel of each module. To store scripts and to transfer
files from the instrument to the host computer, you need a USB flash drive.
•
For information about saving reading buffers to a USB flash drive, see
•
For information about storing and loading scripts to and from a USB flash drive, see
(on page 6-10).
•
For information about file I/O, see
(on page 5-6).
•
For information about saving user setups, see
(on page 2-52).