Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-34
2606B-901-01 Rev. B / May 2018
ConfigPulseIMeasureV()
This KIPulse factory script function configures a current pulse train with a voltage measurement at each point.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
f
,
msg
= ConfigPulseIMeasureV(
smu, bias, level, limit, ton, toff, points, buffer,
tag, sync_in, sync_out, sync_in_timeout, sync_in_abort
)
f
,
msg
= ConfigPulseIMeasureV(
smu, bias, level, limit, ton, toff, points, buffer,
tag, sync_in, sync_out, sync_in_timeout
)
f
,
msg
= ConfigPulseIMeasureV(
smu, bias, level, limit, ton, toff, points, buffer,
tag, sync_in, sync_out
)
f
,
msg
= ConfigPulseIMeasureV(
smu, bias, level, limit, ton, toff, points, buffer,
tag, sync_in
)
f
,
msg
= ConfigPulseIMeasureV(
smu, bias, level, limit, ton, toff, points, buffer,
tag
)
f
A Boolean flag; this flag is
true
when the pulse was successfully configured,
false
when errors were encountered
msg
A string message; if the
f
flag is
false
,
msg
contains an error message; if it is
true
,
msg
contains a string that indicates successful configuration
smu
Instrument channel (for example,
smua
refers to SMU channel A)
bias
Bias level in amperes
level
Pulse level in amperes
limit
Voltage limit (for example, compliance) in volts
ton
Pulse on time in seconds
toff
Pulse off time in seconds
points
Number of pulse-measure cycles
buffer
Reading buffer where pulsed measurements will be stored; if this is
nil
when the
function is called, no measurements will be made when the pulse train is initiated
tag
Numeric identifier to be assigned to the defined pulse train
sync_in
Defines a digital I/O trigger input line; if programmed, the pulse train waits for a
trigger input before executing each pulse
sync_out
Defines a digital I/O trigger output line; if programmed, the pulse train generates a
trigger output immediately before the start of
ton
sync_in_timeout
Specifies the length of time (in seconds) to wait for input trigger; default value is 10 s
sync_in_abort
Specifies whether or not to abort the pulse if an input trigger is not received; if pulse
aborts because of a missed trigger, a timer timeout message is returned;
true
or
false
Details
Data for pulsed voltage measurements are stored in the reading buffer specified by the
buffer
input
parameter.
This function configures a current pulse train with a voltage measurement at each point.
Measurements are made at the end of the
ton
time.