2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-281
Details
These attributes are used to read or write to the operation status SMU X summary registers. Reading
a status register returns a value. The binary equivalent of the returned value indicates which register
bits are set. The least significant bit of the binary number is bit B0, and the most significant bit is bit
B15. The binary equivalent of the value indicates which register bits are set. In the binary equivalent,
the least significant bit is bit B0, and the most significant bit is bit B15. For example, if a value of
1.02
(which is 1,025) is read as the value of the condition register, the binary equivalent is
0000 0100 0000 0010. This value indicates that bit B0 and bit B10 are set.
B15
B14
B13
B12
B11
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
>
>
>
>
>
>
>
>
*
0
0
0
0
0
1
0
0
0
0
0
0
0
0
1
0
* Least significant bit
** Most significant bit
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Enable and transition registers
(on page E-16). The individual bits of this
register are defined in the following table.
Bit
Value
Description
B0
status.operation.instrument.smuX.CALIBRATING
status.operation.instrument.smuX.CAL
Set bit indicates that
smu
X
is unlocked for
calibration.
Bit B0 decimal value: 1
B1-B2
Not used
Not applicable.
B3
status.operation.instrument.smuX.SWEEPING
status.operation.instrument.smuX.SWE
Set bit indicates that
smu
X
is sweeping.
Bit B3 decimal value: 8
B4
status.operation.instrument.smuX.MEASURING
status.operation.instrument.smuX.MEAS
Bit will be set when
taking an overlapped
measurement, but it
will not set when taking
a normal synchronous
measurement.
Bit B4 decimal value:
16
B5-B9
Not used
Not applicable.
B10
status.operation.instrument.smuX.TRIGGER_OVERRUN
status.operation.instrument.smuX.TRGOVR
Set bit indicates an
enabled bit has been
set in the operation
status smu
X
trigger
overrun event register.
Bit B10 decimal value:
1,024
B11-B15
Not used
Not applicable.