Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-342
2606B-901-01 Rev. B / May 2018
This command does not indicate if an overrun occurred in any other part of the trigger model or in any
other trigger object that is monitoring the event. It also is not an indication of an action overrun.
Example
print(trigger.blender[1].overrun)
If an event was ignored, the output is
true
.
If an event was not ignored, the output is
false
.
Also see
(on page 7-342)
trigger.blender[N].reset()
This function resets some of the trigger blender settings to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.blender[
N
].reset()
N
The trigger event blender (up to six)
Details
The
trigger.blender[
N
].reset()
function resets the following attributes to their factory
defaults:
•
trigger.blender[
N
].orenable
•
trigger.blender[
N
].stimulus[
M
]
It also clears
trigger.blender[
N
].overrun
.
Example
trigger.blender[1].reset()
Resets the trigger blender 1 settings to
factory defaults.
Also see
(on page 7-341)
(on page 7-341)
trigger.blender[N].stimulus[M]