Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-282
2606B-901-01 Rev. B / May 2018
As an example, to set bit B0 of the operation status SMU A summary enable register, set
status.operation.instrument.smua.enable
=
status.operation.instrument.smua.C
AL
.
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the
bit to set. To set more than one bit of the register, set
operationRegister
to the sum of their
decimal weights. For example, to set bits B0 and B10, set
operationRegister
to 1,025 (which is
the sum of 1 + 1,024).
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)
Bit
B15
B14
B13
B12
B11
B10
B9
B8
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
32,768
16,384
8,192
4,096
2,048
1,024
512
256
Weights
(2
15
)
(2
14
)
(2
13
)
(2
12
)
(2
11
)
(2
10
)
(2
9
)
(2
8
)
Example
status.operation.instrument.smua.enable =
status.operation.instrument.smua.MEAS
Use a constant to set the MEAS bit
of the operation status SMU A
summary enable register.
Also see
(on page E-7)
status.operation.instrument.smuX.trigger_overrrun.*
(on page 7-282)
status.operation.instrument.smuX.trigger_overrrun.*
This attribute contains the operation status SMU X trigger overrun register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
30 (All bits set)