In addition to the above constants,
questionableRegister
can be set to the numeric equivalent of
the bit to set. To set more than one bit of the register, set
operationRegister
to the sum of their
decimal weights. For example, to set bits B1 and B2, set
questionableRegister
to 6 (which is
the sum of 2 + 4).
Example
status.questionable.over_temperature.enable =
status.questionable.over_temperature.SMUA
Uses a constant to set the SMU A
bit in the questionable status over
temperature summary enable
register.
Also see
(on page 7-303, on page 7-305)
status.questionable.unstable_output.*
This attribute contains the questionable status unstable output summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
6 (All bits set)
Usage
questionableRegister
= status.questionable.unstable_output.condition
questionableRegister
= status.questionable.unstable_output.enable
questionableRegister
= status.questionable.unstable_output.event
questionableRegister
= status.questionable.unstable_output.ntr
questionableRegister
= status.questionable.unstable_output.ptr
status.questionable.unstable_output.enable =
questionableRegister
status.questionable.unstable_output.ntr =
questionableRegister
status.questionable.unstable_output.ptr =
questionableRegister
operationRegister
The status of the questionable status unstable output summary register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other values indicate various bit
settings
Details
These attributes are used to read or write to the questionable status unstable output summary
registers. Reading a status register returns a value. The binary equivalent of the returned value
indicates which register bits are set. The least significant bit of the binary number is bit B0, and the
most significant bit is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Enable and transition registers
(on page E-16). The individual bits of this
register are defined in the following table.