2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-25
Example 1
measurerange = smua.nvbuffer1.measureranges[1]
Store the measure range that was used to make reading number 1.
Example 2
printbuffer(1, 10, smua.nvbuffer1.measureranges)
Print the range values that were used for the first 10 readings saved in dedicated reading buffer 1.
Example output:
1.00000e-07, 1.00000e-07,
1.00000e-07, 1.00000e-07,
1.00000e-07, 1.00000e-07,
1.00000e-07, 1.00000e-07,
1.00000e-07, 1.00000e-07
Also see
(on page 7-23)
(on page 7-25)
(on page 7-26)
(on page 7-27)
(on page 7-29)
(on page 7-30)
(on page 7-31)
bufferVar.n
This attribute contains the number of readings in the buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Clearing the buffer
See
Details
Not applicable
Usage
numberOfReadings
=
bufferVar
.n
numberOfReadings
The number of readings stored in the buffer
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
Details
This read-only attribute contains the number of readings presently stored in the buffer.
For dedicated reading buffers, all buffer attributes are saved to nonvolatile memory only when the
reading buffer is saved to nonvolatile memory.