Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-248
2606B-901-01 Rev. B / May 2018
Trigger event IDs*
Event ID
Event description
trigger.blender[
N
].EVENT_ID
Occurs after a collection of events is detected
trigger.timer[
N
].EVENT_ID
Occurs when a delay expires
trigger.generator[
N
].EVENT_ID
Occurs when the
trigger.generator[
N
].assert()
function is executed
* Use the name of the trigger event ID to set the stimulus value rather than the numeric value. Using the name
makes the code compatible for future upgrades (for example, if the numeric values must change when
enhancements are added to the instrument).
Example
smua.trigger.source.stimulus = digio.trigger[2].EVENT_ID
Configure SMU channel A to start its source action when a trigger event occurs on digital I/O line 2.
Also see
(on page 3-30)
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
This constant contains the source complete event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= smu
X
.trigger.SOURCE_COMPLETE_EVENT_ID
eventID
The source action complete event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.SOURCE_COMPLETE_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
source complete events from this source-measure unit (SMU).
Also see
(on page 3-30)