In this appendix:
Verification ................................................................................ B-1
Adjustment ............................................................................. B-10
Verification
The information in this topic is intended only for qualified service personnel. Some of the
procedures may expose you to hazardous voltages that could result in personal injury or
death. Do not attempt to perform these procedures unless you are qualified to do so.
Use the procedures in this section to verify that the 2606B accuracy is within the limits stated in the
instrument’s one-year accuracy specifications. Perform the verification procedures:
•
When you first receive the instrument to make sure that it was not damaged during shipment.
•
To verify that the instrument meets factory specifications.
•
To determine if calibration is required.
•
After performing a calibration adjustment to make sure the instrument was adjusted properly.
The 2606B contains two independent SMU modules. Each module requires separate verification and
adjustment.
The instructions in this section use the web interface (refer to
(on page 2-18)).
You can also use remote programming commands. Refer to
(on page 5-1) for
information on using commands to control the instrument.
If the instrument is still under warranty and its performance is outside specified limits, contact your
Keithley Instruments representative or the factory to determine the correct course of action.
Appendix B
Calibration