2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-291
Bit
Value
Decimal
value
B0
Not used
Not
applicable
B1
status.operation.instrument.trigger_timer.trigger_overrun.TMR1
2
B2
status.operation.instrument.trigger_timer.trigger_overrun.TMR2
4
B3
status.operation.instrument.trigger_timer.trigger_overrun.TMR3
8
B4
status.operation.instrument.trigger_timer.trigger_overrun.TMR4
16
B5
status.operation.instrument.trigger_timer.trigger_overrun.TMR5
32
B6
status.operation.instrument.trigger_timer.trigger_overrun.TMR6
64
B7
status.operation.instrument.trigger_timer.trigger_overrun.TMR7
128
B8
status.operation.instrument.trigger_timer.trigger_overrun.TMR8
256
B9-B15
Not used
Not
applicable
As an example, to set bit B1 of the operation status trigger timer trigger overrun enable register, set
status.operation.instrument.trigger_timer.trigger_overrun.enable =
status.operation.instrument.trigger_timer.trigger_overrun.TMR1
.
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the
bit to set. To set more than one bit of the register, set
operationRegister
to the sum of their
decimal weights. For example, to set bits B1 and B4, set
operationRegister
to 18 (which is the
sum of 2 + 16).
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)
Bit
B15
B14
B13
B12
B11
B10
B9
B8
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
32,768
16,384
8,192
4,096
2,048
1,024
512
256
Weights
(2
15
)
(2
14
)
(2
13
)
(2
12
)
(2
11
)
(2
10
)
(2
9
)
(2
8
)
Example
status.operation.instrument.trigger_timer.trigger_overrun.enable =
status.operation.instrument.trigger_timer.trigger_overrun.TMR3
Uses a
constant to
set the timer
3 bit of the
operation
status trigger
timer overrun
enable
register.
Also see
(on page E-7)
status.operation.instrument.trigger_timer.*