2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-301
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)
Bit
B15
B14
B13
B12
B11
B10
B9
B8
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
32,768
16,384
8,192
4,096
2,048
1,024
512
256
Weights
(2
15
)
(2
14
)
(2
13
)
(2
12
)
(2
11
)
(2
10
)
(2
9
)
(2
8
)
Example
operationRegister = status.operation.trigger_overrun.SMUA +
status.operation.trigger_overrun.TRGBLND
status.operation.trigger_overrun.enable = operationRegister
Uses constants to set bit B1 and bit B10 of the operation status trigger overrun summary enable register.
Also see
(on page E-7)
(on page 7-266)
status.operation.user.*
These attributes manage the operation status user register set of the status model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (RW)
Yes
Status reset
Not saved
0
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
32,767 (All bits set)
Usage
operationRegister
= status.operation.user.condition
operationRegister
= status.operation.user.enable
operationRegister
= status.operation.user.event
operationRegister
= status.operation.user.ntr
operationRegister
= status.operation.user.ptr
status.operation.user.condition =
operationRegister
status.operation.user.enable =
operationRegister
status.operation.user.ntr =
operationRegister
status.operation.user.ptr =
operationRegister
operationRegister
The status of the operation status user register; a zero (0) indicates no bits set
(also send 0 to clear all bits); other values indicate various bit settings