Bit
Value
Decimal value
B0
Not used
Not applicable
B1
status.operation.instrument.digio.trigger_overrun.LINE1
2
B2
status.operation.instrument.digio.trigger_overrun.LINE2
4
B3
status.operation.instrument.digio.trigger_overrun.LINE3
8
B4
status.operation.instrument.digio.trigger_overrun.LINE4
16
B5
status.operation.instrument.digio.trigger_overrun.LINE5
32
B6
status.operation.instrument.digio.trigger_overrun.LINE6
64
B7
status.operation.instrument.digio.trigger_overrun.LINE7
128
B8
status.operation.instrument.digio.trigger_overrun.LINE8
256
B9
status.operation.instrument.digio.trigger_overrun.LINE9
512
B10
status.operation.instrument.digio.trigger_overrun.LINE10
1,024
B11
status.operation.instrument.digio.trigger_overrun.LINE11
2,048
B12
status.operation.instrument.digio.trigger_overrun.LINE12
4,096
B13
status.operation.instrument.digio.trigger_overrun.LINE13
8,192
B14
status.operation.instrument.digio.trigger_overrun.LINE14
16,384
B15
Not used
Not applicable
As an example, to set bit B1 of the operation status digital I/O overrun enable register, set
status.operation.instrument.digio.trigger_overrun.enable =
status.operation.instrument.digio.trigger_overrun.LINE1
.
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the
bit to set. To set more than one bit of the register, set
operationRegister
to the sum of their
decimal weights. For example, to set bits B1 and B10, set
operationRegister
to 1,026 (which is
the sum of 2 + 1,024).