Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-156
2606B-901-01 Rev. B / May 2018
Example
PulseIMeasureV(smua, 0.001, 1.0,
20e-3, 40e-3, 10)
SMU A outputs 1 mA and dwells for 40 ms, outputs
1 A and dwells for 20 ms. The voltage measurements
occur during each 20 ms dwell period. After the
measurement, the output returns to 1 mA and dwells
for 40 ms. This pulse-measure process repeats nine
more times.
Also see
(on page 5-18)
PulseVMeasureI()
This KIPulse factory script function performs a specified number of pulse V, measure I cycles.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
PulseVMeasureI(
smu, bias, level, ton, toff, points
)
smu
Instrument channel (for example,
smua
refers to SMU channel A)
bias
Bias level in volts
level
Pulse level in volts
ton
Pulse on time in seconds
toff
Pulse off time in seconds
points
Number of pulse-measure cycles
Details
If any parameters are omitted or
nil
, the operator is prompted to enter them using the front panel.
Data for pulsed current measurements, voltage levels, and timestamps are stored in
smu
X
.nvbuffer1
.
To perform the specified number of pulse V, measure I cycles, this function:
1. Sets the
smu
to output
bias
volts and dwell for
toff
seconds
2. Sets the
smu
to output
level
volts and dwell for
ton
seconds
3. Performs voltage measurement with source at
level
volts
4. Sets the
smu
to output
bias
volts for
toff
seconds
5. Repeats steps 2 through 4 for the remaining
points
pulse-measure cycles
Figure 122: PulseVMeasureI()