Section 5: Remote commands
2606B System SourceMeter® Instrument Reference Manual
5-18
2606B-901-01 Rev. B / May 2018
KIPulse factory script
The KIPulse factory script provides examples of how to generate pulses and to provide a simple
pulsing interface. Pulses can be generated using the functions listed below.
Please note the following information about the KIPulse factory script:
This factory script only operates on the channels present in the instrument executing the pulse
functions. These functions will not operate correctly if you attempt to access instrument channels
over the TSP-Link
®
interface.
The KIPulse factory scripts are general purpose examples that may not be suitable for all use
cases. Very short pulses (less than 1 ms pulse width) may require optimization of the examples
provided by the factory script in order to achieve settled measurements.
The
PulseIMeasureV()
and
PulseVMeasureI()
functions may be accessed from the front
panel. The remaining functions may only be accessed remotely.
KIPulse tag parameter pulse functions
(on page 5-19) to configure a pulse train
and assign the configuration to the
tag
parameter (use
QueryPulseConfig()
to inspect
configured pulse trains). Use the initiation
InitiatePulseTest()
function to execute the pulse
trains assigned to its
tag
arguments. The conditions listed in the table below must be true for these
functions to execute successfully.
Conditions that must be true for successful function execution
Conditions for Config
functions:
Conditions for
InitiatePulseTest
functions
Conditions for
InitiatePulseTestDual
functions
Source autorange (I and V)
off
Output on
Output on
Measure autorange (I and
V) off
There is enough free space
in the buffer
There is enough free space in
the buffer
Measure NPLC < ton
Buffer append mode is on
when pulse train is >1 point
Buffer append mode is on
when pulse train is >1 point
Measure autozero OFF or
ONCE
Separate unique
source-measure units (SMUs)
for each tag
Same NPLC setting for each
tag
Same toff for each tag
Use the
KIPulse simple pulse functions
(on page 5-19) to specify and perform a specified number of
pulse-measure cycles.
The following functions make up the KIPulse factory script.