2606B System SourceMeter® Instrument Reference Manual
Section 5: Remote commands
2606B-901-01 Rev. B / May 2018
5-19
KIPulse tag parameter pulse functions
(on page 7-34)
(on page 7-40)
ConfigPulseIMeasureVSweepLin()
ConfigPulseVMeasureISweepLin()
ConfigPulseIMeasureVSweepLog()
(on page 7-38)
ConfigPulseVMeasureISweepLog()
(on page 7-44)
(on page 7-105)
(on page 7-157)
KIPulse simple pulse functions
(on page 7-155)
(on page 7-156)
Advanced features for KIPulse tag parameter pulse functions
Variable off time between pulses in a pulse train
The KIPulse “Configure” functions will accept the
toff
parameter as a table, or as a number. The
table allows you to define different off times to be used after each pulse. The following should be
noted:
•
If
toff
is passed as a number or only a single value is used in the table, it will be used for all
points in a multiple point pulse.
•
The number of times specified in the table must match the number of points called for in the
sweep.
•
The times used in tables must match for dual channel pulsing.
•
Each specified off time must adhere to the duty cycle limits for the specified pulsing region.
Simultaneous IV measurement during pulse
The KIPulse “Configure” functions will optionally accept an extra reading buffer to activate
simultaneous IV measurements during pulsing. Previous usage of passing in a reading buffer or a nil
(for no measurement) is still supported.