2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-295
status.operation.measuring.*
This attribute contains the operation status measuring summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
6 (All bits set)
Usage
operationRegister
= status.operation.measuring.condition
operationRegister
= status.operation.measuring.enable
operationRegister
= status.operation.measuring.event
operationRegister
= status.operation.measuring.ntr
operationRegister
= status.operation.measuring.ptr
status.operation.measuring.enable =
operationRegister
status.operation.measuring.ntr =
operationRegister
status.operation.measuring.ptr =
operationRegister
operationRegister
The status of the operation status measuring summary register; a zero (0) indicates
no bits set (also send 0 to clear all bits); other values indicate various bit settings
Details
These attributes are used to read or write to the operation status measuring summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Enable and transition registers
(on page E-16). The individual bits of this
register are defined in the following table.
Bit
Value
Description
B0
Not used
Not applicable.
B1
status.operation.measuring.SMUA
Bit will be set when SMU A is taking an
overlapped measurement, but it will not be
set when taking a normal synchronous
measurement.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2
status.operation.measuring.SMUB
Bit will be set when SMU B is taking an
overlapped measurement, but it will not be
set when taking a normal synchronous
measurement.
Bit B2 decimal value: 4
Binary value: 0000 0100
B3-B15
Not used
Not applicable.