Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-104
2606B-901-01 Rev. B / May 2018
Example
smua.reset()
smua.source.rangev = 5
smua.source.rangei = 1
smua.source.levelv = 0
smua.measure.rangev = 5
smua.measure.rangei = 1
smua.measure.nplc = 0.01
smua.measure.autozero = smua.AUTOZERO_ONCE
smua.nvbuffer1.clear()
smua.nvbuffer1.appendmode = 1
smua.source.output = smua.OUTPUT_ON
f1, msg1 = ConfigPulseVMeasureI(smua, 0, 5, 1, 0.002, 0.2, 10, smua.nvbuffer1, 1)
if f1 == true then
f2, msg2 = InitiatePulseTest(1)
print("Initiate message:", msg2)
else
print("Config errors:", msg1)
end
Configure channel A to generate a pulse train. If no errors are encountered, initiate the pulse train. Channel A
pulses voltage from a bias level of 0 V to a pulse level of 5 V. The pulse level is present for 2 ms and the bias
level for 200 ms, with a 1 A limit setting. A total of 10 pulses is generated, and the measurement data is stored
in
smua.nvbuffer1
. This pulse train is assigned to
tag = 1
.
Also see
(on page 7-34)
ConfigPulseIMeasureVSweepLin()
ConfigPulseIMeasureVSweepLog()
(on page 7-38)
(on page 7-40)
ConfigPulseVMeasureISweepLin()
ConfigPulseVMeasureISweepLog()
(on page 7-44)
(on page 5-18)